The Eclipse 850 Electrocardiograph provides features and performance typically found on units costing much more. A large preview display saves time and money by reducing retakes. Battery operation gives the unit enhanced mobility. Computerized interpretation, using the well-respected Glasgow interpretive program, provides a silent second opinion with documented accuracy. Additional options for internal storage, fax and communications, and exercise stress, allow you to configure a system that meets your needs for today and tomorrow.
Features:
Click here to see the actual product literature with large photos!
Specifications:
| Dimensions: | 11.75" x 18.16" x 4.14" (299mm x 462mm x 105mm) |
| Weight: | 14.2 lbs (6.45 Kg) |
| Display: | Backlit 320 x 240 LCD |
| Keyboard: | Full alphanumeric keypad |
| Input/Output: | Compact flash (proprietary true IDE) |
| Units w/ communications: Serial RS-232 (9 pin D) | |
| Telephone line interface (RJ11C) | |
| Analog / TTL / ExTOL output (25 pin D) | |
| Data Storage: | Optional -- 40 patient records |
| Power Req.: | AC operation: 115/230V AC +/- 10%, 0.80/0.40 A, 50/60 Hz |
| Fuses: for 115V, 1.0A, 250V type T; for 230V, 0.5A 250V type T | |
| Battery operation: 16.8V NiCad rechargeable battery pack | |
| (80 minutes w/o printing or 40 minutes w/ 40 ECGs) | |
| Printout: | Device: 216mm thermal dot array |
| Paper: 8.5" x 11" (US letter) or 210mm x 300 mm (A4) | |
| Paper type: thermal sensitive | |
| Chart speeds: 10, 25, 50 mm/sec | |
| Gain: 5, 10, 20 mm/mV; 10 mm/mV with 5 mm/mV chest; | |
| 20 mm/mV with 10 mm/mV chest | |
| Formats: standard or cabrera 3, 4, 6, or 12 channel; add. rhythm | |
| Acquisition: | Lead selection: I, II, III, aVR, aVF, V1, V2, V3, V4, V5, V6, V4F |
| Supports Frank X, Y, Z and Nehb D, A, J | |
| Interpretation opt.: Diagnosis, measurements, reason statments | |
| Modes: Automatic, rhythm, or manual | |
| Frequency response: Meets or exceeds ANSI/AAMI EC11-1991 | |
| Input impedance: Meets or exceeds ANSI/AAMI EC11-1991 | |
| Electrode offset tolerance: +/- 300 mV | |
| A/D conversion: 5uV LSB | |
| Artifact filter response: 40Hz, -3db | |
| Data resolution: 500 samples/second, 5uV | |
| Safety: | Leakage current: patient < 10uA, chassis < 100uA |
| Defibrillation protection: To 5000V, 400J | |
| Warranty: | Three years parts and labor |
| Due to continuing product innovations, these specifications are subject to change without notice. Contact us for details. |
Ordering Options:
Eclipse 850 non-interpretive electrocardiograph
with Optional Storage
with Communication configuration
with Exercise Stress configuration
Eclipse 850 interpretive electrocardiograph
with Optional Storage
with Communication configuration
with Exercise Stress configuration
Other options:
Eclipse cart
Light-duty cart
Astro-Trace® universal clips
CardioSens® ULTRA II disposable electrodes
Assurance 50® fade resistant paper